What is the optimal needle penetration of the OPTIC?

Solution

In general the optimum needle penetration for OPTIC (and LINEX) is 43 mm measured from the top of the septum nut.

The injection speed depends a bit on the syringe type, injection volume and temperature of injection. But in general a fast injection (50-200µl/sec) is good with OPTIC.

 

 
Was this article helpful? yes / no