Is it possible to make an EGA thermogram with OPTIC?

Solution

EGA (Evolved Gas Analysis) is possible with the OPTIC, just set a low ramp rate in the method and start with a low temperature and end with the wanted temperature.

To set a very low ramp rate the expert method should be used, in this method it is possible to set a ramp rate of 0.1°C/sec. Make sure the End Time is long enough.

 

 